660°C Working Standard PRTs
Model WSP660 PRTs use our special wire wound RTD elements assembled into inconel probes. They have similar accuracy and long-term stability as SSP670 secondary standard PRTs. They can be used for calibration of furnaces, sand bath and other calibration work. Available in standard models and custom design.
| Specifications
|
How
to order
|
| Resistance at 0°C:
|
|
|
100Ω or custom design
|
| Temperature range:
|
|
|
-250 - 660°C
|
| Temperature coefficient:
|
|
|
0.003850 Ω/Ω/°C
or 0.003925 Ω/Ω/°C
minimum
|
| Stability:
|
|
|
R0 typical drift < 0.02°C
after 500 hours at 660°C
|
| Thermal shock:
|
|
|
R0 typical drift < 0.005°C after
100 times from 25 to 660°C
|
| Dimensions:
|
|
|
0.188" and 0.25" dia. x 12"
L standard, custom sizes available
|
|
WSP660 Model number
| A
Tolerance
at 0°C:
|
|
|
A
=±0.06%, IEC DIN class A
B
=±0.12%, IEC DIN class B
|
| 120
Probe
length:
|
|
|
120 = 12.0 inches
Specify in 0.5" increment
|
| N4
Number of leads:
|
|
|
N4 = 4 leads
|
| L36
Lead
wire length in inches
|
|
|
L36 = 36 inches
|
| Sample part number:
|
|
|
WSP660A120N4L36
|
|
|
|
|
|
|
|
|
|