Test Results
of ASP Products
|
1111The
following test data is base on the test results of ASP and ASP¡¯s
customers. The products include Model PRP300, PRP500, PRP 660,
PRP850, WSP300, WSP500, WSP660, WSP850, SSP500, SSP670, and SSP850.
|
|
|

|
|
1.Long
term stability test
|
|
Model
|
T (¡ãC)
|
Time (Hours)
|
Test DR0.01 (¡ãC)
|
Specification (¡ãC)
|
|
PRP300
|
300
|
1,000
|
0.001
|
0.002
|
|
PRP500
|
500
|
1,000
|
0.008
|
0.015
|
|
PRP660
|
660
|
1,000
|
0.012
|
0.030
|
|
PRP850
|
850
|
1,000
|
0.030
|
0.050
|
|
WSP300
|
250
|
1,000
|
0.0006
|
0.001
|
|
WSP500
|
350
|
500
|
0.001
|
0.002
|
|
WSP500
|
500
|
500
|
0.005
|
0.008
|
|
WSP660
|
660
|
500
|
0.008
|
0.020
|
|
WSP850
|
850
|
500
|
0.015
|
0.030
|
|
SSP500
|
500
|
500
|
0.004
|
0.006
|
|
SSP670
|
660
|
500
|
0.006
|
0.010
|
|
SSP850
|
850
|
500
|
0.012
|
0.015
|
|
|
|
2.Hysteresis:
|
|
(1)Test result: 0.0005¡ãC Max.
from ¨C196 to 660¡ãC.
11 ASP specification:
0.001¡ãC Max. from ¨C196 to 660¡ãC.
|
|
(2)Test result: 0.001¡ãC Max.
from ¨C196 to 960¡ãC.
11ASP
specification: 0.002¡ãC Max. from ¨C196 to 960¡ãC.
|
|
|
|
|

|
|

|
|
|
|
ASP
products are widely used in industrial process control and laboratory
metrology.
|
|
|

|
|
|
|
Excellent long term stability and wide
temperature range is the specialty of ASP RTDs and PRTs.
|
|
|

|
|
|
|
ASP accepts major credit cards order.
|
|
|

|
|